1. Product advantages:
(1) Equipped with both digital and analog signal outputs (±2.4V), it is compatible with liquid chromatography systems from different manufacturers.
(2) 1-7 levels of signal gain adjustment, allowing response tuning based on sample concentration.
(3) Up to 100 groups of method storage space; stored methods can be directly recalled for pharmacopoeia-specified items.
(4) The instrument panel features real-time chromatogram display for monitoring data acquisition, facilitating preliminary experiments.
- Comprehensive Upgrade in Reliability & Safety
(1) The instrument performs multiple self-checks upon power-on, with error correction control. Real-time monitoring and alarms are available for temperature, pressure, flow rate and other parameters. Sound and light alarms will be triggered under abnormal conditions. The multi-layer safety protection design simplifies maintenance, reduces maintenance costs and extends service life.
(2) Automatic switching between standby and working modes effectively lowers power consumption.
(3) Password-protected screen lock to prevent unauthorized operation and error log display to ensure GLP compliance.
- Intelligent Operation for Modern Application
Full reverse control is realized for core parameters including temperature, flow rate and signal gain. Real-time feedback of operating status timely reflects instrument conditions to ensure stable and smooth experimental operation.
With decades of expertise in evaporative light scattering detection, our Company provides comprehensive pre-sales and after-sales services. Our professional application team supports customized method development and technical analysis. Regular application training sessions are offered on a regular basis. In addition, we provide long-term preferential supply of spare parts, consumables and reagents.
2. Key Parameters
| Type |
Low-temperature |
| Light Source |
Laser |
| Detection Element |
PMT |
| Signal Output |
Digital & Analog signals |
| Evaporation Temperature Range |
R.T.–130°C |
| Gas Source |
Nitrogen or Air (1 L/min–4 L/min) |
| Liquid Flow Rate Range |
10 μL/min–3 mL/min |
| Nebulizer Heating (Optional) |
R.T.–85°C |
| Signal Gain |
Multi-level adjustable gain |
3. Applications
Medicine/Pharmaceutic

Food Industry

GB / T35867-2018/ISO11707:2009 Analysis of phospholipid in Lecithin by HPLC-ELSD

Analysis of lactose and sucrose in infant food and dairy products
Chemical industry field

Sample: 25ul Shodex Pullulan P-5 25mg/ml
Other details:
- Transportation:By air if not specified.