Home > Products > Detector> UV/Vis on-column detector

UV/Vis on-column detector

Overview of TriSep ™ –2100 UV/Vis On-column Detector

Unimicro's TriSep ™ UV/Vis detector is made for on-column detection technologies, specially for pressurized Capillary Electro-Chromatography (pCEC) systems, Capillary Electrophoresis (CE) system and micro-HPLC. It is a high-performance and multi-functional ultra-violet visible spectrophotometric on-column detector, incorporating both deuterium and tungsten halogen lamps. Each unit has 3 measurement modes - single wavelength, dual wavelength and wavelength scanning.

Features of TriSep™ –2100 UV/Vis On-column Detector

  • Excellent S/N ratio performance

    An improved optical system and a high-order digital filter have enhanced the unit's Signal to Noise ratio. Enhanced wavelength accuracy, reproducibility, and broader wavelength range measurement have heightened the unit's basic performance.

  • Advanced functions

    The dual wavelength mode provides chromatograms or ratio chromatograms of two wavelengths simultaneously, and the wavelength-scanning mode provides absorbance spectra. There is also a full set of Time Program functions.

  • No optical alignment needed when lamps are replaced

    Adjustment of the optical alignment is not required when the lamps are replaced. The deuterium lamp has a longer life than previous models. And the unit records the total time that the lamp has been used.

Specifications of TriSep™ –2100 UV/Vis On-column Detector

Item

TriSep™—2100 detector

Light source

Deuterium lamp

Wavelength range

190-600nm Cut-off filter to eliminate second order diffraction is automatically activated for the 371—600nm range.

Spectral bandwidth

8nm

Wavelength accuracy

1nm

Wavelength reproducibility

0.1nm

Drift

5×10-4 AU/hour Max.

Noise Level

6×10-6 AU

Dual wavelength mode

Measurement Wavelength

Selectable, two wavelengths in 190—370nm or 371—600nm range

Sampling frequency

1.2 sec for one wavelength

Response

Selectable in 10 steps corresponding to time constant 0.05, 0.1, 0.5, 1.0, 1.5, 2.0, 3.0, 6.0, 8.0, 10.0 seconds

Range

Can be set between 0.0001 and 2.56 AUFS in 0.0001 AUFS steps

Zero adjustment

Auto zero function, baseline shift function

Polarity switching Wavelength scanning function

Possible Performs wavelength scanning with flow stopped. Data stored in three files, one of which stores background. Background is subtracted from sample scans. File data is not saved up (lost when power is turned OFF).

Wavelength steps

1-5nm, selectable in 5 steps,

Scanning speed

2-5nm when W lamp is used. 10-50nm/sec, settable in 5 steps (according to the wavelength step)

Spectrum plot output speed

1, 3, 10 nm/sec

Ratio chromatogram

Outputs the absorption ratio of two wavelengths

Time Program

Available in detector or by system controller

 

Set items

Wavelength (including dual wavelengths), auto zero, range, marker, response, wavelength scanning, event, polarity, lamp ON/OFF, loop

Number of steps (detector program)

Maximum 32 steps

Output

Output for recorder

10mV recorder terminals

Output for integrator

6 steps: 0.5, 1, 2, 4, 1.25 and 2.5 AU/V Integrator output terminals output chromatogram in single wavelength mode or the Ch1 chromatogram in dual wavelength mode. Ratio chromatogram, spectrum, and Ch2 chromatogram are output from the recorder output terminals.

Lamp hour-meter

Records up to 9999.9 hours

Size

260(W) 140(H) 430(D) mm

Weight

13 kg

Operating temperature range

4°C — 35°C

Power supply

110-120V AC, 10 % 150VA 50-60Hz
220-230V AC/240V AC, 10 % 150VA 50-60Hz