BiSep™-1100 High Performance Liquid Chromatography System

BiSep™-1100 High Performance Liquid Chromatography System (HPLC), manufactured by Unimicro Technologies, Inc., is a newly released separation system with good quality, reliability and accuracy.

BiSep™-1100 HPLC System is composed of detector and pumps. With modular design, the system is easily converted to microbore HPLC, which can save 75% of solvent and sample.

BiSep™-1100 Detector

Overview

The Unimicro Technologies BiSep detector is high-performance, multi-function system ultra-violet with a visible spectrophotometric detector, for use in high performance liquid chromatography. The BiSep detector uses a deuterium lamp as its light source and is intended primarily for analyses in the ultraviolet region. The BiSep detector incorporates both deuterium and tungsten halogen lamps. The deuterium lamp is used for ultraviolet applications, as in the BiSep detector, while the tungsten lamp extends the analytical capability into the visible region. Each unit has 3 measurement modes - single wavelength, dual wavelength and wavelength scanning. The dual wavelength mode performs simultaneous detection of two wavelengths, and can provide chromatograms of both the wavelengths, or a chromatogram of one wavelength and one ratio chromatogram. In wavelength scanning mode, the absorbance spectrum is measured. The wavelength scanning mode is designed to be used while the flow is stopped.

Features
  • Excellent S/N ratio performance

    An improved optical system and a high-order digital filter have enhanced the unit's Signal to Noise ratio. Enhanced wavelength accuracy, reproducibility, and broader wavelength range measurement have heightened the unit's basic performance.

  • Advanced functions

    The dual wavelength mode provides chromatograms or ratio chromatograms of two wavelengths simultaneously, and the wavelength scanning mode provides absorbance spectra. There is also a full set of Time Program functions.

  • No optical alignment needed when lamps are replaced

    No adjustment of optical alignment is required when the lamps are replaced. The deuterium lamp has a longer life than previously, and the unit records the total time the lamp has been used.

Specifications

Item

BiSep ™ —1100 detector

Light source

Deuterium lamp

Wavelength range

190-600nm Cut-off filter to eliminate second order diffraction is automatically activated for the 371—600nm range.

Spectral bandwidth

8nm

Wavelength accuracy

1nm

Wavelength reproducibility

0.1nm (*1)

Drift

2×10-4 AU/hour Max. (250nm, room temperature constant, air in the cell) 3-10-4 AU/hour Max. (250 nm, room temperature fluctuation less than 2°C, air in the cell)

Noise Level

0.35×10-5 AU

Cell path length

8 µL

Cell pressure tolerance

12MPa {120kgf/cm2 }

Cell materials in contact with liquid

SUS316, teflon, quartz

Cell inlet, outlet tubing diameter

SUS316 tubing 1/ 16” OD 0.3mmID

Dual wavelength mode (*3)

Measurement Wavelength

Selectable, two wavelengths in 190—370nm or 371—600nm range

Sampling frequency

1.2 sec for one wavelength

Response

Selectable in 10 steps corresponding to time constant 0.05, 0.1, 0.5, 1.0, 1.5, 2.0, 3.0, 6.0, 8.0, 10.0 seconds

Range

Can be set between 0.0001 and 2.56 AUFS in 0.0001 AUFS steps

*1 Reproducibility when the wavelength is changed after turning the power ON in single wavelength mode.
*2 In single wavelength mode.
*3 Noise level in dual wavelength mode varies depending on the wavelength. It may be necessary to use less sensitive detector settings (higher Range values) when performing dual wavelength analysis.

Item

BiSep ™ -1100 detector

Zero adjustment

Auto zero function, baseline shift function

Polarity switching Wavelength scanning function

Possible Performs wavelength scanning with flow stopped. Data stored in three files, one of which stores background. Background is subtracted from sample scans. File data is not saved up (lost when power is turned OFF).

 

Wavelength steps

1-5nm, selectable in 5 steps, 2-5nm when W lamp is used. 10-50nm/sec, settable in 5 steps (according to the wavelength step) 1, 3, 10 nm/sec

Scanning speed

Spectrum plot output speed

Ratio chromatogram

Outputs the absorption ratio of two wavelengths

Time Program

Available in detector or by system controller

 

Set items

Wavelength (including dual wavelengths), auto zero, range, marker, response, wavelength scanning, event, polarity, lamp ON/OFF, loop

Number of steps (detector program)

Maximum 32 steps

Output

 

Output for recorder

10mV recorder terminals

Output for

integrator

6 steps: 0.5, 1, 2, 4, 1.25 and 2.5 AU/V Integrator output terminals output chromatogram in single wavelength mode or the Ch1 chromatogram in dual wavelength mode. Ratio chromatogram, spectrum, and Ch2 chromatogram are output from the recorder output terminals.

Lamp hour-meter

Records up to 9999.9 hours

Size

260(W) 140(H) 430(D) mm

Weight

13 kg

Operating temperature range

4°C — 35°C

Power supply

110-120V AC, 10 % 150VA 50-60Hz
220-230V AC/240V AC, 10 % 150VA 50-60Hz

BiSep™-1100 Pump

Overview

The pump is a dual reciprocating plunger solvent delivery module. It offers improved accuracy and sensitivity in the BiSep™-1100system. In addition to the delivery module, the BiSep™-1100 system also requires, at a minimum, an injector, column, detector, high voltage module and system controller. Note that each of these is purchased separately from the pumping unit. Please consult our neareat branch/sales office or agent regarding the components that are necessary for your system.

Features

  • High Gradient Accuracy and Ultra-Low Flow Pulsation

    Both the duration and the magnitude of the pump’s flow pulsation are significantly reduced in comparison to conventional pumps. This is enabled by a design which drives the plunger at high speed and reduces discharge to a micro volume: 10µl per plunger stroke. This lack of flow pulsationcapacity to a micro column 10 allows detectors such as differential refractometers and electrochemical detectors to be used in their high sensitivity range, an application impossible with conventional pumps. The solvent delivery system also provides highly accurate gradent delivery in a semi-micro volume range.

  • A Range of Gradient Delivery Functions

    The pump can be used in both low- and high- pressure gradient delivery modes. The high-pressure gradient delivery mode is highly accurate with minimum time lag.

  • Automatic Plunger Seal Rinsing

    An optional automatic mechanism rinses the back of the plunger seal, precenting premature
    seal failure due to crystallization of high-salt-concentration buffer solutions.

Specifications

The system specifications

Item

Specification

Pump type

Micro- volum double plunger pump(approx. 10 µ L per stroke )

Pumping methods

Constant flow delivery and constant pressure delivery

 

 

Constant flow pumping

Flow rate setting range

1.0-39.2MPa (10-400kgf/cm2 )0.001-5mL/min[1.0-39.2MPa(10-400kgf/cm2 )]
1.0-19.6MPa(10-200kgf/cm2 )5.001-9.999mL/min[1.0-19.6MPa(10-200kgf/cm2 )]

Flow rate
accuracy

± 2% or ± 2 µ L/min, whichever is greater(when water is pumped at 0.01-5mL/min, 1.0-39.2MPa(10-400kgf/cm2 ), with room temperature 20°C)
A different accuracy applies to low-pressure gradient delivery

Flow rate
precision

Within ± 0.3% (when water is pumped at 0.1-5mL/min, 1.0-39.2MPa(10-400kgf/cm2 ), with room temperature 20°C) )

 

Constant pressure pumping

Pressure setting range

1.0-39.2MPa(10-400kgf/cm2)[0.1MPa (1kgf/cm2 )steps]

Pressure
accuracy

± 10% or1.0MPa (10kgf/cm2 ), whichever is greater

Pressure limits

Upper and lower limits

Damper

Interior capacity 400 µ L approx

Liquid-contacting part materials

SUS316, ruby, sapphire, Teflon ®, HastelloyC®, PEEK, perfluoroelastomer
Zirconia, PCTFE

Suction filter

10 µ m mesh

Line filter

5 µ m mesh, capacity 30 µ L approx.

Time program

Commands for flow rate, pressure, EVENT functions, LOOP (for program repetition ) 10files,320 steps.

Pressure display accuracy

­­­­ ± 2% or 0.5MPa (5kgf/cm2 ), whichever is greater

Plunger cleaning line

Automatic rinsing possible with optional kit

Size

W260 × H 140 × D 420mm , excluding projections

Weight

11kg

Operating temperature range

4-35°C

High-pressure gradient system specifications

Item

Specification

Number of solvents mixed

2

Gradient profile

Step and linear possible at multiple levels

Maximum program steps

Max.320 steps per program. 20program files

Program duration

Max.999.99 minutes (in 0.01minute steps

Mix ratio setting range

0-100% (in 0.1% steps)

Concentration accuracy

± 1% [with aqueous acetone/water, at 0-100%, 0.1-3mL/min,1.0-39.2MPa(10-400kgf/cm2)]

Flow rates possible

0.001-9.999mL/min

Copyright 2002-2005 Unimicro Technologies. All Rights Reserved