Vitamin B

Experimental Section
Instruments
Experiments were performed on a TriSep™-2010 Capillary Electrochromatography system (Shanghai Unimicro Analytical Technologies Co., Ltd.) with a 0-30 kV high voltage power supply, a high speed scanning UV detector, a solvent delivery system and HW system software.
Reagent: methanol(HPLC grade);deionized water;TEA; phosphoric acid;Sodium dihydrogen phosphate;Thiamine hydrochloride (VB1); Pyridoxine HCl (VB6); Niacinamide;
Conditions
Column:Capillary C 18
Mobile phase:A: Methanol
B: 10mM NaH2PO4 buffer, 0.3%TEA, pH=2.3(H3PO4)
A: B = 10 : 90
Flow:0.2 ml/min
Wavelength:270 nm
Temperature:Room temperature
Results


Figure 1 Capillary Electrochromatogram of vitamin,-3.0 kV/-19.3μA
1. VB
1 2. Niacinamide 3. VB6

 